Model Number 45007
Device Problems Complete Blockage (1094); Suction Problem (2170)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/16/2023
Event Type  malfunction  
Event Description
It was reported that the jetstream catheter was clogged.A 2.1mm jetstream xc atherectomy catheter was selected to treat peripheral arterial disease of the right lower extremity.During the procedure, while inside the patient, after about 5 minutes of treatment time it was noted that the suction was not working as well due to thrombus likely lodged in the aspiration port.The catheter was removed from body and re-primed, however, the clog did not clear.The procedure was completed successfully using another jetstream catheter.No patient complications were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
JETSTREAM XC
Type of Device
CATHETER, PERIPHERAL, ATHERECTOMY
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
model farm road
cork T12 Y K88
EI   T12 YK88
Manufacturer Contact
rachel shields
4100 hamline ave n
arden hills, MN 55112
6512422111
MDR Report Key17710992
MDR Text Key322958221
Report Number2124215-2023-48969
Device Sequence Number1
Product Code MCW
UDI-Device Identifier08714729889892
UDI-Public08714729889892
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K133023
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 09/08/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/08/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number45007
Device Catalogue Number45007
Device Lot Number0031871698
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/16/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/22/2023
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
-
-