Model Number AED PRO
Device Problems Self-Activation or Keying (1557); Failure of Device to Self-Test (2937)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
Complainant alleged that during functional testing, the device prompted a "unit failed" and "shock button stuck"messages.Complainant indicated that there was no patient involvement in the reported malfunction.
 
Manufacturer Narrative
Zoll medical corporation has not received the device for evaluation and this complaint is still under investigation.
 
Manufacturer Narrative
The device was returned to zoll medical corporation for evaluation.The customer's report of "self test failed" was observed and attributed to a stuck pin on the flex cable.The flex cable was replaced as a precaution.The device was recertified and returned to the customer.The customer's report of "critical errors observed in log" observed during review of the device data logs.The device was put through extensive testing without duplicating the report.The shock button gasket, shock switch retainer and shock button switch collar were replaced as a pre-caution.Analysis of reports of this type has not identified an increase in trend.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AED PRO
Type of Device
DEFIBRILLATOR
Manufacturer (Section D)
ZOLL MEDICAL CORPORATION
269 mill road
chelmsford MA 01824
Manufacturer Contact
scott august
269 mill road
chelmsford, MA 01824
9784219552
MDR Report Key17713087
MDR Text Key322991357
Report Number1220908-2023-03470
Device Sequence Number1
Product Code MKJ
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
P160022
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 08/18/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/08/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberAED PRO
Device Catalogue NumberAED PRO
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received08/18/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/01/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-