• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COCHLEAR BONE ANCHORED SOLUTIONS AB NUCLEUS CI632 PROFILE PLUS WITH SLIM MODIOLAR ELECTRODE; NUCLEUS 24 COCHLEAR IMPLANT SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COCHLEAR BONE ANCHORED SOLUTIONS AB NUCLEUS CI632 PROFILE PLUS WITH SLIM MODIOLAR ELECTRODE; NUCLEUS 24 COCHLEAR IMPLANT SYSTEM Back to Search Results
Model Number CI632
Device Problem Migration (4003)
Patient Problems Vomiting (2144); Dizziness (2194); Fluid Discharge (2686)
Event Date 03/17/2023
Event Type  Injury  
Event Description
Per the clinic, the patient post-operative experienced a severe gusher, and then severe vomiting and dizziness.Ct scan revealed the implant had migrated into the auditory canal and was resting on the vestibular nerve.The device was explanted (date unknown) and was re-implanted with a new device in the same procedure.
 
Manufacturer Narrative
This report is submitted on september 12, 2023.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NUCLEUS CI632 PROFILE PLUS WITH SLIM MODIOLAR ELECTRODE
Type of Device
NUCLEUS 24 COCHLEAR IMPLANT SYSTEM
Manufacturer (Section D)
COCHLEAR BONE ANCHORED SOLUTIONS AB
konstruktionsvigen 14
molnlycke, 435 3 3
AS  435 33
Manufacturer (Section G)
COCHLEAR BONE ANCHORED SOLUTIONS AB
konstruktionsvigen 14
molnlycke, 435 3 3
SW   435 33
Manufacturer Contact
roberta loveday
1 university avenue
macquarie university, nsw 2109
AS   2109
MDR Report Key17726252
MDR Text Key323147165
Report Number6000034-2023-03106
Device Sequence Number1
Product Code MCM
UDI-Device Identifier09321502036580
UDI-Public(01)09321502036580(11)221013(17)241012
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970051
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 08/23/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/11/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Model NumberCI632
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/23/2023
Was Device Evaluated by Manufacturer? No
Date Device Manufactured10/13/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age67 YR
Patient SexFemale
-
-