Catalog Number PIP-20
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Implant Pain (4561)
Event Date 08/14/2023
Event Type  Injury  
Event Description
Revision surgery - due to pain.
 
Manufacturer Narrative
Additional reporting on this event will be provided as a supplemental report to this document as soon as it becomes available.
 
Manufacturer Narrative
The reason for this revision surgery, the agent reported "(implant removal due to breakage and localised response.Implanted by mr.Rep present during implantation.No issues implanting implant.Good post-op result.Patient re-presented with redness and pain.Samples suggest no presence of infection.Implant removed due to continued pain last week, signs of bone erosion, 2-stage-revision.No other implant replaced so far.Silicone bridge reported as separated.Samples taken.No growth as yet for infection.Per required on implant.Implant available from (non-decontaminated at present).Hospital was requested to decontaminate implant and return to osteotec for quarantine)." the actual length of in-vivo for the item(s) listed is unknown as the original surgery date or dtickets were not provided to determine it.This evaluation is limited in scope as the item(s) associated with this investigation was not returned to djo surgical - austin for examination.The surgery was completed as intended and without incident.If additional information regarding the reported event is submitted at a future date, this investigation will be re-evaluated.There was no information submitted with this complaint about any patient activities, accidents, or medical contraindications that may have contributed to the reported event.There were no findings during this evaluation that indicate that the reported device(s) was defective.The surgeon performed this procedure to remedy the patient's condition.No further action is deemed necessary.A review of the device history record(s) show that the reported component(s) used in the previous surgery, when released for use, met design and manufacturing requirements.There were no nonconforming material reports associated with the product(s) that may have contributed to the reported event.The device(s) was verified to have gone through an acceptable sterilization process and was within its expiration date at the time of the previous surgery.Customer complaint history of the reported device(s) showed no present trends or on-going issues that are needing a review.The root cause of this complaint was a revision surgery implant removal due to breakage and localised response.No information was submitted with the complaint regarding pre-existing conditions of the patient or any activities the patient was involved in that may have contributed to the event.There are multiple factors that may contribute to an event that are outside of the control of djo surgical.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
20 SILICONE PIP IMPLANT (STERILE PACKED)
Type of Device
PROSTHESIS, FINGER, CONSTRAINED, POLYMER
Manufacturer (Section D)
ENCORE MEDICAL L.P
9800 metric blvd
austin TX 78758
Manufacturer (Section G)
ENCORE MEDICAL L.P
9800 metric blvd
austin TX 78758
Manufacturer Contact
james mcmahon
9800 metric blvd
austin, TX 78758
MDR Report Key17726685
MDR Text Key323146549
Report Number1644408-2023-01233
Device Sequence Number1
Product Code KYJ
Combination Product (y/n)N
Reporter Country CodeUK
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 11/28/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/11/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date04/30/2024
Device Catalogue NumberPIP-20
Device Lot Number13560D
Was Device Available for Evaluation? No
Date Manufacturer Received11/09/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/30/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-