ST. JUDE MEDICAL, INC. AMPERE¿ RF ABLATION GENERATOR; CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FLUTTER
|
Back to Search Results |
|
Model Number H700489 |
Device Problem
Failure to Deliver Energy (1211)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 08/31/2023 |
Event Type
malfunction
|
Event Description
|
During a supraventricular tachycardia ablation procedure, output issues resulted in unsuccessful troubleshooting measures and the procedure was cancelled.The ampere rf generator would not deliver any power to the ablation catheter.The rf patch connection was checked and replaced, and the rf connection cable was checked and replaced with no resolve.The ablation was unable to be performed and the procedure was not completed.
|
|
Manufacturer Narrative
|
The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be determined.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.No non-conformances associated with the reported event were identified.No capa is required; the root cause of the reported issue could not be confirmed to be a manufacturing, design, or quality system related occurrence.The reported issue will continue to be monitored for trends and reviewed in qdr (quality data review).
|
|