Model Number PM1152 |
Device Problems
Failure to Interrogate (1332); No Pacing (3268)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 09/05/2023 |
Event Type
Injury
|
Event Description
|
During an in clinic follow up, the device was unable to be interrogated using inductive telemetry and a loss of pacing was observed.The device was explanted and replaced to resolve the event.The patient was stable.
|
|
Manufacturer Narrative
|
The reported events of no pacing and no inductive telemetry were confirmed.The device was received with no telemetry communication and no output.Visual inspection of the header attachment area detected an anomaly between the epoxy header and titanium case.The device was cut open to enable further testing and the battery was found in normal range.Feedthrough leak test was performed, indicating a device hermeticity breach.This is consistent with feedthrough damage as a result of fluid intrusion between the header and case, and subsequent fluid ingress to internal electronics.Hybrid circuitry was tested, and the results indicated normal current drain.A manufacturing process anomaly may have occurred, which resulted in the header bonding anomaly.The device is included in the assurity and endurity pacemakers header anomaly advisory issued by abbott on 15 march 2021.
|
|