It was reported that during initial implant procedure, patient's new right ventricular (rv) lead exhibited difficulty in advancing through patient's right ventricular septum.It was further noted that the lead was over torqued, and the lead body surface was wavy.The lead was not installed and the procedure was completed using an alternate lead on (b)(6) 2023.There were no patient consequences.
|
The reported event of lead body over-torqued/twisted was not confirmed.As received, a complete lead with a stylet was returned in one piece for analysis.Visual inspection and x-ray examination of the lead found the helix was stretched, bent, and clogged with blood/tissue.The helix mechanism/extension length test couldn¿t perform due to the damaged helix consistent with procedural damage.Electrical testing did not find any indication of conductor fractures or internal shorts.No anomalies were found.
|