• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: HOYA CORPORATION PENTAX TOKYO OFFICE PENTAX; CAMERA HEAD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HOYA CORPORATION PENTAX TOKYO OFFICE PENTAX; CAMERA HEAD Back to Search Results
Model Number PVK-J10
Device Problem Optical Obstruction (3002)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/30/2023
Event Type  malfunction  
Event Description
The time of event is unknown.There was no report of patient harm.Video image failure (cloudy).
 
Manufacturer Narrative
The product was returned to pentax medical for repair.Our technician checked the returned unit and confirmed that the ccd module with drive pcb cloudy (not clear view).Based on the result, we concluded that it was caused due to the moisture condensation in the ccd module with drive pcb.Based on the technical report ""hr-rpt-0586(image failure)"" and/or the risk analysis results, it was evaluated to submit mdr.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENTAX
Type of Device
CAMERA HEAD
Manufacturer (Section D)
HOYA CORPORATION PENTAX TOKYO OFFICE
tsutsujigaoka 1-1-110
akishima-shi, tokyo 196-0 012
JA  196-0012
Manufacturer (Section G)
HOYA CORPORATION PENTAX TOKYO OFFICE
tsutsujigaoka 1-1-110
akishima-shi, tokyo 196-0 012
JA   196-0012
Manufacturer Contact
william goeller
3 paragon drive
montvale, NJ 07645
8004315800
MDR Report Key17734362
MDR Text Key323256552
Report Number9610877-2023-58857
Device Sequence Number1
Product Code FEA
Combination Product (y/n)N
Reporter Country CodeSF
PMA/PMN Number
K220465
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Distributor
Reporter Occupation Other
Type of Report Initial
Report Date 09/13/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/13/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberPVK-J10
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/30/2023
Date Manufacturer Received08/30/2023
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-