• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH RESTORATION ADM X3 INS 28/52; PROSTHESIS, HIP, SEMI-CONSTRAINED, UNCEMENTED, METAL/POLYMER, NON-POROUS, CALICU

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH RESTORATION ADM X3 INS 28/52; PROSTHESIS, HIP, SEMI-CONSTRAINED, UNCEMENTED, METAL/POLYMER, NON-POROUS, CALICU Back to Search Results
Catalog Number 7236-2-852
Device Problem Detachment of Device or Device Component (2907)
Patient Problems Bone Fracture(s) (1870); Joint Dislocation (2374); Inadequate Osseointegration (2646)
Event Date 08/25/2023
Event Type  Injury  
Manufacturer Narrative
An event regarding disassociation involving an adm liner was reported.The event was not confirmed.Method & results: -product evaluation and results: material analysis, visual, functional and dimensional inspections could not be performed as the device was not returned.-clinician review: no medical records were received for review with a clinical consultant.-product history review: review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.-complaint history review: there have been no other similar events for the lot referenced.Conclusions: it was reported that the patient was revised due to disassociation of the ceramic head from the adm liner.Loosening of the stem, periprosthetic fracture around the acetabular shell, and movement of the shell were also reported.The exact cause of the event could not be determined because insufficient information was provided.Further information such as return of the device, pathology reports, pre- and post-operative x-rays and the revision operative report as well as patient history and follow-up notes are needed to complete the investigation for determining root cause.No further investigation for this event is possible at this time.If devices and/or additional information become available to indicate further evaluation is warranted, this record will be reopened.
 
Event Description
As reported: "please find attached information from index and revision surgeries of a right total hip.The revision was secondary to dislocation.It was discovered that the ceramic head disassociated from the poly insert.Also discovered was a crack in the acetabulum and the appearance that the cup had moved.It was also determined that the stem was loose.All implants were removed.A multihole cup was implanted with multiple screws and an mdm liner.The femur was prepared for an insignia stem and trialed.The hip remained grossly unstable so it was decided to leave the patient with a girdlestone.The plan being to come back after the cup has healed and place a rest mod with constrained liner.No further information available from the doctor or hospital.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESTORATION ADM X3 INS 28/52
Type of Device
PROSTHESIS, HIP, SEMI-CONSTRAINED, UNCEMENTED, METAL/POLYMER, NON-POROUS, CALICU
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-CORK
ida industrial estate
carrigtwohill NA
EI   NA
Manufacturer Contact
brad curtis
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key17735200
MDR Text Key323249990
Report Number0002249697-2023-01026
Device Sequence Number1
Product Code MEH
UDI-Device Identifier07613327315967
UDI-Public07613327315967
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K182468
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Health Professional
Reporter Occupation Physician
Type of Report Initial
Report Date 09/13/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/13/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue Number7236-2-852
Device Lot Number10139551
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/25/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/27/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age80 YR
Patient SexMale
-
-