• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DENTSPLY LLC AUTOMATRIX SHAFT; INSTRUMENTS, DENTAL HAND

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DENTSPLY LLC AUTOMATRIX SHAFT; INSTRUMENTS, DENTAL HAND Back to Search Results
Catalog Number 62422603
Device Problem Break (1069)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
While no serious injury resulted in this event, if this malfunction recurred, it could cause or contribute to a serious injury or require medical or surgical intervention to preclude such.This event, therefore, is reportable per 21cfr part 803.The device is available for evaluation, though has not been returned as of this report.Evaluation results will be submitted as they become available.
 
Event Description
In this event it is reported that a automatrix shaft broke during use.No injury occurred.
 
Manufacturer Narrative
Investigation result: 9-28-2023: returned product was 1 flexshaft date code 0622 with coil deformation/weld failure causing the product to not function properly.Capa opened to address weld failures for product manufactured by sarasota since december 2020 (date code 1220) through implementation date of august 2022 (0822).The batch information provided in this case is inaccurate based on the date code of the returned product.Item# 62422603 batch# 06611021 was kitted in jan 2023, and the component item# 24703 batch# 06298365 was also produced in jan 2023 and would have a date coded of 0123.Complaint is considered substantiated however due to the product returned meeting criteria per capa.(nwv).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AUTOMATRIX SHAFT
Type of Device
INSTRUMENTS, DENTAL HAND
Manufacturer (Section D)
DENTSPLY LLC
38 west clarke avenue
milford DE 19963
Manufacturer (Section G)
DENTSPLY LLC
38 west clarke avenue
milford DE 19963
Manufacturer Contact
hannah seevaratnam
221 west philadelphia st.
york, PA 17401
7178457511
MDR Report Key17735258
MDR Text Key323252483
Report Number2515379-2023-00099
Device Sequence Number1
Product Code DZN
UDI-Device IdentifierD010624226011
UDI-PublicD010624226011
Combination Product (y/n)N
Reporter Country CodeGM
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Dentist
Type of Report Initial,Followup
Report Date 09/29/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/13/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue Number62422603
Device Lot Number06611021
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Distributor Facility Aware Date09/12/2023
Date Manufacturer Received09/12/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Type of Device Usage A
Patient Sequence Number1
-
-