• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MICROPORT CRM S.R.L. GALI; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MICROPORT CRM S.R.L. GALI; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO Back to Search Results
Model Number GALI 4LV SONR CRT-D 2844
Device Problems Display or Visual Feedback Problem (1184); Application Program Problem: Parameter Calculation Error (1449)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/16/2023
Event Type  malfunction  
Event Description
Reportedly, on a remote monitoring report pre-irm, it is noted an atrial lead higher than 3000 ohms.The atrial lead impedance post-irm, the value was within a normal range (500 ohms).Is it a display issue?.
 
Manufacturer Narrative
The device involved in this mdr report is not approved in the united states; however, it is similar to a device manufactured by microport crm that was cleared or approved by fda for marketing in the united states.
 
Event Description
Reportedly, on a remote monitoring report pre-irm, it is noted an atrial lead higher than 3000 ohms.The atrial lead impedance post-irm, the value was within a normal range (500 ohms).
 
Manufacturer Narrative
The issue reported in this complaint is related to a programmer bug.The lead impedance value is saved in the device when the mri starts.The bug occurs when the programmer managed the statistic decoding.No issue is suspected on the device.For more details, please refer to the attached analysis report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GALI
Type of Device
DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO
Manufacturer (Section D)
MICROPORT CRM S.R.L.
via crescentino s.n
saluggia (vc) 13040
IT  13040
Manufacturer (Section G)
MICROPORT CRM S.R.L.
via crescentino s.n
saluggia (vc) 13040
IT   13040
Manufacturer Contact
elodie vincent
via crescentino s.n
saluggia (vc) 13040
IT   13040
MDR Report Key17735829
MDR Text Key323400771
Report Number1000165971-2023-00702
Device Sequence Number1
Product Code NIK
Combination Product (y/n)N
Reporter Country CodeFR
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 11/05/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/13/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberGALI 4LV SONR CRT-D 2844
Device Catalogue NumberGALI 4LV SONR CRT-D 2844
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received10/11/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient EthnicityNon Hispanic
-
-