Model Number GALI 4LV SONR CRT-D 2844 |
Device Problems
Display or Visual Feedback Problem (1184); Application Program Problem: Parameter Calculation Error (1449)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 08/16/2023 |
Event Type
malfunction
|
Event Description
|
Reportedly, on a remote monitoring report pre-irm, it is noted an atrial lead higher than 3000 ohms.The atrial lead impedance post-irm, the value was within a normal range (500 ohms).Is it a display issue?.
|
|
Manufacturer Narrative
|
The device involved in this mdr report is not approved in the united states; however, it is similar to a device manufactured by microport crm that was cleared or approved by fda for marketing in the united states.
|
|
Event Description
|
Reportedly, on a remote monitoring report pre-irm, it is noted an atrial lead higher than 3000 ohms.The atrial lead impedance post-irm, the value was within a normal range (500 ohms).
|
|
Manufacturer Narrative
|
The issue reported in this complaint is related to a programmer bug.The lead impedance value is saved in the device when the mri starts.The bug occurs when the programmer managed the statistic decoding.No issue is suspected on the device.For more details, please refer to the attached analysis report.
|
|