Model Number H700977 |
Device Problem
Failure to Power Up (1476)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 07/13/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
The results of the investigation are inconclusive since the device was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.No non-conformances associated with the reported event were identified.Based on the information received, the cause of the reported power issue and subsequent cancellation remain unknown.
|
|
Event Description
|
During the procedure with the patient prepped, the workmate claris display workstation continuously booted up and the case was abandoned.
|
|
Manufacturer Narrative
|
Additional information: d9, g1, g3, h2, h3, h6 one claris computer sn (b)(6) was returned for investigation.Visual inspection revealed the unit was damaged being repair and multiple boards and pieces were falling off.Due to the state of the return, no functional testing was possible.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.No non-conformances associated with the reported event were identified.
|
|