• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC. WORKMATE¿ CLARIS¿ COMPUTER; COMPUTER, DIAGNOSTIC, PROGRAMMABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC. WORKMATE¿ CLARIS¿ COMPUTER; COMPUTER, DIAGNOSTIC, PROGRAMMABLE Back to Search Results
Model Number H700977
Device Problem Failure to Power Up (1476)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/13/2023
Event Type  malfunction  
Manufacturer Narrative
The results of the investigation are inconclusive since the device was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.No non-conformances associated with the reported event were identified.Based on the information received, the cause of the reported power issue and subsequent cancellation remain unknown.
 
Event Description
During the procedure with the patient prepped, the workmate claris display workstation continuously booted up and the case was abandoned.
 
Manufacturer Narrative
Additional information: d9, g1, g3, h2, h3, h6 one claris computer sn (b)(6) was returned for investigation.Visual inspection revealed the unit was damaged being repair and multiple boards and pieces were falling off.Due to the state of the return, no functional testing was possible.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.No non-conformances associated with the reported event were identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WORKMATE¿ CLARIS¿ COMPUTER
Type of Device
COMPUTER, DIAGNOSTIC, PROGRAMMABLE
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.
one st. jude medical drive
st. paul MN 55117
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.
one st. jude medical drive
st. paul MN 55117
Manufacturer Contact
janna parks
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key17737147
MDR Text Key323317716
Report Number2184149-2023-00174
Device Sequence Number1
Product Code DQK
UDI-Device Identifier05415067009777
UDI-Public05415067009777
Combination Product (y/n)N
Reporter Country CodeIN
PMA/PMN Number
K151911
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 02/06/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/13/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberH700977
Device Lot Number6012319
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received01/30/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/14/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-