During ivtm therapy, the thermogard ivtm system (sn (b)(6)) reaches the target temperature for the patient with little or significant delays.The target temperature was 34,5 degree celsius at maximum cooling mode, and within 3 hours, the patient temperature was not under 34,77 degree celsius.Patient's status information was requested but the customer did not provide a response.
|
The customer reported a complaint that "the thermogard ivtm system (sn (b)(6)) reaches the target temperature for the patient with little or significant delays" was confirmed based on the review of the patient data log but not during the functional testing.The root cause of the reported issue could not be conclusively determined as the reported issue was not duplicated during the functional testing, and the thermogard console functioned as intended.During visual inspection, no physical damage was observed on the thermogard console.During the event log review, no significant discrepancies were noted.However, the patient data review noted that the patient was not cooled, thus confirming the customer's complaint.The functional test was performed, and no issue was observed.The console passed all tests, and the temperature reading inputs were correct.No device malfunction was noted.
|