• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: KARL STORZ SE & CO. KG ENDOFLATOR 50

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

KARL STORZ SE & CO. KG ENDOFLATOR 50 Back to Search Results
Model Number UI500
Device Problem Device Alarm System (1012)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/15/2020
Event Type  malfunction  
Manufacturer Narrative
The previous reportability decision for this event was reversed based on a retrospective review.The reported device was returned and evaluation was performed.The reported complaint was confirmed.The evaluator found the pressure sensor board and the filter board boards were defective.The boards were replaced.The reported issue will continue to be tracked and trended.The event is filed under internal (b)(6) complaint id: (b)(4).
 
Event Description
It was reported that the endoflator 50 stopped working and gave an error during lap chole surgery.Per report, the system gave the error "3ff: power on test error - a system failure occurred.You have limited access to the settings menu only, restart the device in order to proceed, contact (b)(6) if problem persists".The surgeon used a back-up unit to complete the surgery.The device was reported to be in excellent condition prior to use.There was no report of injury to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOFLATOR 50
Type of Device
ENDOFLATOR 50
Manufacturer (Section D)
KARL STORZ SE & CO. KG
dr.-karl-storz-strasse 34
tuttlingen, 78532
GM  78532
Manufacturer (Section G)
KARL STORZ SE & CO. KG
dr.-karl-storz-strasse 34
tuttlingen, 78532
GM   78532
Manufacturer Contact
anja fair
2151 e grand ave.
el segundo, CA 90245
4242188247
MDR Report Key17741340
MDR Text Key323369041
Report Number9610617-2023-00237
Device Sequence Number1
Product Code FCX
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K161554
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 09/13/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/13/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberUI500
Device Catalogue NumberUI500
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/01/2022
Date Manufacturer Received02/16/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/01/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-