The lot was manufactured from january 17, 2023 to january 20, 2023.The device was received for evaluation.Visual inspection was performed and a small surface cut was observed on a segment of the tubing.The reported condition was verified.Examination on the cut mark suggested the tubing may have inadvertently come in contact with a sharp object such as a needle during use, which resulted a small surface cut.During manufacturing, the unit was leak tested and therefore, the surface cut on the tubing would not have passed manufacturing leak test which suggested the surface cut may not have occurred during manufacturing.The cause of the condition may be attributed to use error.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|