ORTHO-CLINICAL DIAGNOSTICS, INC. VITROS CHEMISTRY PRODUCTS PHYT SLIDES; IN-VITRO DIAGNOSTICS Back to Search Results
Catalog Number 8298671
Device Problem Low Test Results (2458)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/13/2023
Event Type  malfunction  
Manufacturer Narrative
The investigation determined that lower than expected vitros phenytoin (phyt) results were obtained from a single level non-vitros mas quality controls, lot ocr25113, using two lots of vitros phyt tested on vitros xt7600 integrated system, j76000432.The assignable cause of the event is unknown.Quality control results for both lots of vitros phyt showed inaccuracy and imprecision for the two levels at higher concentrations.Although acceptable performance was obtained using an alternate non-vitros biorad quality control, this control was not run consistently and therefore it could not be confirmed that the issue was isolated to the mas control.Acceptable results were obtained at various times on all levels with the mas control with no action to the analyzer, therefore a reagent issue with either of the two lots is not a likely contributor to the event.Additionally, diagnostic precision testing performed using vitros alkp and vitros phyt lot 2625-0185-2783 were within ortho acceptable guidelines, confirming that the vitros xt7600 system and the vitros phyt reagent were not likely contributors to the event.The customer indicated that they are following the instructions for use for calibration preparation and quality control fluid handling, therefore a fluid handling issue cannot be confirmed as a contributor to the event.Continual tracking and trending of complaints has not identified any signals that would point to a potential systemic issue with vitros phyt slide lots 2625-0185-2295 or 2625-0185-2783.
 
Event Description
A customer contacted the ortho clinical diagnostics (ortho) technical solution center (tsc) to report lower than expected vitros phenytoin (phyt) results were obtained from a single level non-vitros mas quality controls, lot ocr25113, using two lots of vitros phyt tested on vitros xt7600 integrated system, j76000432.Phyt lot 2625-0185-2295: vitros phyt lot 2625-0185-2295 mas fluid results of 22.40, 21.19 and 20.71 ug/ml versus the mas peer mean of 28.7 ug/ml phyt lot 2625-0185-2783: vitros phyt lot 2625-0185-2783 mas fluid results of 18.09, 17.81, 18.94, 19.68, 20.75, and 22.36 ug/ml versus the mas peer mean of 28.7 ug/ml biased results of the magnitude and direction observed may lead to inappropriate physician action if they were to occur undetected on patient samples.The lower than expected results were obtained from non-patient quality control fluids.There was no allegation of patient harm as a result of this event.This report is number seven of nine mdr¿s for this event.Nine 3500a forms are being submitted for this event as nine devices were involved.This report corresponds to ortho clinical diagnostics inc (ortho) complaint numbers (b)(4), and reportability assessment (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VITROS CHEMISTRY PRODUCTS PHYT SLIDES
Type of Device
IN-VITRO DIAGNOSTICS
Manufacturer (Section D)
ORTHO-CLINICAL DIAGNOSTICS, INC.
100 indigo creek drive
rochester NY 14626
Manufacturer (Section G)
ORTHO-CLINICAL DIAGNOSTICS, INC.
513 technology boulevard
rochester NY 14626
Manufacturer Contact
laurie o'riordan
microtyping systems
1295 southwest 29th avenue
pompano beach, FL 33069
9549709500
MDR Report Key17748251
MDR Text Key323698136
Report Number0001319809-2023-00101
Device Sequence Number1
Product Code DIP
UDI-Device Identifier10758750004690
UDI-Public10758750004690
Combination Product (y/n)N
Reporter Country CodeSP
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 09/13/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/14/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/15/2023
Device Catalogue Number8298671
Device Lot Number2625-0185-2783
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/28/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/19/2023
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-