OLYMPUS WINTER & IBE GMBH TELESCOPE "TRUEVIEW II", 4 MM, 30°, AUTOCLAVABLE; ARTHROSCOPE
|
Back to Search Results |
|
Model Number A70941A |
Device Problems
Break (1069); Melted (1385)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 08/18/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
The device was returned to olympus for evaluation and the customer's allegation was confirmed.The evaluation found lens damage, light guide damage, burning/ melting of gold painting on the outer tube and poor lighting due to a broken light guide.The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation.
|
|
Event Description
|
The customer reported to olympus, the telescope "trueview ii", has lens damage.The defect was found during reprocessing.The device was returned for evaluation.During the device evaluation, the following reportable malfunctions were found: lens damage, light guide damage, and burning/ melting of gold painting on the outer tube.There were no reports of patient harm.This medical device report (mdr) is being submitted to capture the reportable malfunctions reported by the customer and found during evaluation.
|
|
Manufacturer Narrative
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the root cause of the event could not be determined although it can be presumed it was due to user error, improper handling as well as application of excessive force.Olympus will continue to monitor field performance for this device.
|
|
Search Alerts/Recalls
|
|
|