OLYMPUS WINTER & IBE GMBH TELESCOPE "TRUEVIEW II", 4 MM, 30°, AUTOCLAVABLE; ARTHROSCOPE Back to Search Results
Model Number A70941A
Device Problems Break (1069); Melted (1385)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/18/2023
Event Type  malfunction  
Manufacturer Narrative
The device was returned to olympus for evaluation and the customer's allegation was confirmed.The evaluation found lens damage, light guide damage, burning/ melting of gold painting on the outer tube and poor lighting due to a broken light guide.The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation.
 
Event Description
The customer reported to olympus, the telescope "trueview ii", has lens damage.The defect was found during reprocessing.The device was returned for evaluation.During the device evaluation, the following reportable malfunctions were found: lens damage, light guide damage, and burning/ melting of gold painting on the outer tube.There were no reports of patient harm.This medical device report (mdr) is being submitted to capture the reportable malfunctions reported by the customer and found during evaluation.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the root cause of the event could not be determined although it can be presumed it was due to user error, improper handling as well as application of excessive force.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TELESCOPE "TRUEVIEW II", 4 MM, 30°, AUTOCLAVABLE
Type of Device
ARTHROSCOPE
Manufacturer (Section D)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg, deutschland 22045
GM  22045
Manufacturer (Section G)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg, deutschland
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key17752319
MDR Text Key323701966
Report Number9610773-2023-02587
Device Sequence Number1
Product Code HRX
UDI-Device Identifier04042761010542
UDI-Public04042761010542
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K950076
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 12/22/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/14/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberA70941A
Device Lot Number611038
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/21/2023
Was the Report Sent to FDA? No
Date Manufacturer Received12/20/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/06/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-