• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STERIS MEXICO, S. DE R.L. DE C.V. V-PRO 1 PLUS STERILIZER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STERIS MEXICO, S. DE R.L. DE C.V. V-PRO 1 PLUS STERILIZER Back to Search Results
Device Problem Use of Device Problem (1670)
Patient Problem Burn(s) (1757)
Event Date 08/18/2023
Event Type  malfunction  
Manufacturer Narrative
A steris service technician arrived onsite following the reported event to inspect the v-pro 1 plus sterilizer and found the unit to be operating properly.No repairs were required and the unit was returned to service.The technician was informed that the employee was not wearing proper ppe, specifically gloves as stated in the operator manual.The v-pro 1 plus sterilizer operator manual states (6-11), "steris recommends (in accordance with ansi/aami st58, 2005) wearing chemical-resistant gloves when using the sterilization unit." the operator manual further states (1-2), "danger - chemical injury hazard: any visible liquids in the chamber must be treated as concentrated hydrogen peroxide.Observe all hydrogen peroxide handling precautions.When handling hydrogen peroxide, wear appropriate personal protective equipment." additionally, user facility personnel should ensure all instruments are properly dry prior to placement in the v-pro 1 plus sterilizer.The v-pro 1 plus operator manual states (a-1), "a.Dry all items thoroughly.B.Ensure all moisture is removed from all internal parts (including lumens).C.Only dry items are to be placed in sterilization unit." the technician counseled the user facility on the importance of wearing proper ppe, specifically gloves, while operating their v-pro 1 plus sterilizer.No additional issues have been reported.
 
Event Description
The user facility reported an employee experienced a burn on their arm while handling items that were processed in their v-pro 1 plus sterilizer.The employee sought medical treatment at the facility's er and returned to work.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
V-PRO 1 PLUS STERILIZER
Type of Device
STERILIZER
Manufacturer (Section D)
STERIS MEXICO, S. DE R.L. DE C.V.
avenida avante 790
parque industrial guadalupe
guadalupe, 67190
MX  67190
Manufacturer (Section G)
STERIS MEXICO, S. DE R.L. DE C.V.
avenida avante 790
parque industrial guadalupe
guadalupe, 67190
MX   67190
Manufacturer Contact
daniel davy
5960 heisley road
mentor, OH 44060
MDR Report Key17759533
MDR Text Key323543471
Report Number3005899764-2023-00047
Device Sequence Number1
Product Code MLR
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 09/15/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/15/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/18/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/02/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-