The investigation determined that a higher than expected sodium result was obtained from a college of american pathologists (cap).Proficiency sample repeat processed using vitros chemistry products sodium (na+) slides, lot 4211-1096-5681 on a vitros xt 7600 integrated system.The assignable cause of the event cannot be determined.An issue with the age of the cap sample cannot be ruled out as a contributor to the event.Per cap, opened specimens are stable at 2-8 degrees celsius for 7 days.The customer initially processed the sample in june 2023 and repeated the sample in august 2023, obtaining the higher than expected result above.Historical quality control results for vitros na+ lot 4211-1096-5681 showed some imprecision at higher concentrations, however current performance was within expectation, indicating that a vitros na+ lot 4211-1096-5681 performance issue is not a likely contributor to the event.Continual tracking and trending of complaints has not identified any signals that would point to a potential systemic issue with vitros na+ lot 4211-1096-5681.As precision testing was not run at the time of the event, an instrument related issue could not be confirmed nor ruled out as a contributing factor of the events.
|