PHYSIO-CONTROL, INC. - 3015876 LIFEPAK(R) 15 MONITOR/DEFIBRILLATOR; AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE) Back to Search Results
Model Number 15
Device Problem Failure to Sense (1559)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/19/2023
Event Type  malfunction  
Event Description
The customer contacted stryker to report that their device failed to sense shockable rhythm during intervention on a patient.In this state, the device may not be able to provide defibrillation therapy, if it were needed.The patient associated with the reported event did not survive.The health care provider confirmed that the device use did not contribute to the patient outcome.
 
Manufacturer Narrative
Stryker contacted the customer to request additional information on the patient.The customer provided stryker with the available patient information.Patient fields in which information is not provided were intentionally left blank.   a stryker service representative performed an initial evaluation of the customer¿s device and was not able to duplicate the reported issue.Proper device operation was observed through functional and performance testing.The device was subsequently returned to the customer.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFEPAK(R) 15 MONITOR/DEFIBRILLATOR
Type of Device
AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE)
Manufacturer (Section D)
PHYSIO-CONTROL, INC. - 3015876
11811 willows road ne
redmond WA 98052
Manufacturer (Section G)
PHYSIO-CONTROL, INC. - 3015876
11811 willows road ne
redmond WA 98052
Manufacturer Contact
dara friedman
11811 willows road ne
redmond, WA 98052
4258674000
MDR Report Key17760641
MDR Text Key324081844
Report Number0003015876-2023-01669
Device Sequence Number1
Product Code MKJ
UDI-Device Identifier00883873871690
UDI-Public00883873871690
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P160026
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 09/15/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/15/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number15
Device Catalogue Number99577-001957
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/22/2023
Date Manufacturer Received08/20/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/11/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Age54 YR
Patient SexMale
Patient Weight100 KG
-
-