Device evaluated by manufacturer and h6.Evaluation codes: email is: (b)(6) two (2) units were received without their original packaging, in used condition, and decontaminated.Per visual inspection, the proximal end female luer connectors were observed to be broken at the tip.Although both units arrived with the threads of the luer connectors damaged, the functionality of the returned units was able to be tested by a leak test.Both units failed the leak test.The complaint was confirmed.After reviewing the mitigations that are performed during the manufacturing process to detect damaged component, the root cause is that the connectors became damaged after the product left the manufacturing facilities as this damaged would be detected during manufacturing process.A device history record (dhr) review reported no discrepancies or non-conformances during the manufacturing of the reported lot number.This failure will continue to be monitored.
|