C.R. BARD, INC. (COVINGTON) -1018233 NICORE NITINOL GUIDEWIRE Back to Search Results
Catalog Number 150NSS35
Device Problem Material Fragmentation (1261)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/25/2023
Event Type  malfunction  
Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event Description
It was reported that the guidewire was broken when pulling out.Per follow-up information received from ibc on 06sep2023, stated that there was no broken piece was found in patient body.
 
Manufacturer Narrative
The reported event is confirmed cause unknown.Visual evaluation of the returned photo sample noted one opened (without original packaging), used nitinol guidewire.Visual inspection of the photo sample noted the tip of the guidewire is piercing out of the coating.This does not meet specification per inspection procedure which states "no components are to be missing, incorrect, or damaged.A potential root cause for this failure mode could be ¿material selection, device design, degradation¿.A review of the device history record did not show any problems or conditions that would have contributed to the reported issue.The instructions for use were found adequate and state the following: "warning: inspect all guidewires for damage prior to use.Bending or kinking during or prior to placement could damage the guidewire.Do not attempt to use the guidewire if it has been damaged.Use of a damaged wire may result in damage to the urinary tract.Do not reshape the guidewire by any means.Attempting to reshape the guidewire may cause damage resulting in release of fragments into the urinary tract.Failure to exercise proper caution may result in damage to the urinary tract.Do not manipulate or remove the guidewire through a metal cannula or needle.This may result in destruction/separation of the outer jacket of the wire requiring retrieval.Failure to comply with the warnings could result in damage to the guidewire to include, but not limited to: wire breakage, abrasion of the coating, release of guidewire fragments into the urinary system, all of which might require intervention." correction: e.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : the actual/suspected device was inspected.
 
Event Description
It was reported that the guidewire was broken when pulling out.Per follow-up information received from ibc on 06sep2023, stated that there was no broken piece was found in patient body.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NICORE NITINOL GUIDEWIRE
Type of Device
NICORE NITINOL GUIDEWIRE
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer (Section G)
C.R. BARD INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer Contact
xeeroy rada
8195 industrial blvd
covington 30014
7707846100
MDR Report Key17763854
MDR Text Key323588967
Report Number1018233-2023-06838
Device Sequence Number1
Product Code EZB
UDI-Device Identifier00801741051524
UDI-Public(01)00801741051524
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 12/20/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/18/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number150NSS35
Device Lot NumberNGHP1214
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/29/2023
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/21/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/11/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-