• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER SURGICAL, INC. ZIMMER AIR DERMATOME; DERMATOME HANDPIECE, PNEUMATIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER SURGICAL, INC. ZIMMER AIR DERMATOME; DERMATOME HANDPIECE, PNEUMATIC Back to Search Results
Model Number N/A
Device Problems Loose or Intermittent Connection (1371); Physical Resistance/Sticking (4012)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/14/2023
Event Type  malfunction  
Event Description
It was reported that during testing that the depth of cut lever was not working correctly; not graduating properly and sticky.There was no harm or injury as there was no patient involvement.Also, an alternate device was available for immediate use.Due diligence is complete and no additional information is available.No adverse event reported.
 
Manufacturer Narrative
This event is recorded by zimmer biomet under (b)(4).G2, country event occurred in: australia the product is in the process of being evaluated by zimmer biomet.Once the product investigation has been completed, a follow up/final report will be submitted.
 
Manufacturer Narrative
This event is recorded by zimmer biomet under (b)(4).Review of the most recent repair record determined the thickness control lever was loose.The lever, ball plunger, semi-circle bearings, and vespel bearings were replaced and resolved the reported issue.Review of the device history record identified no deviations or anomalies during manufacturing.A definitive root cause cannot be determined.The event is confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
No additional information is available regarding the incident.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ZIMMER AIR DERMATOME
Type of Device
DERMATOME HANDPIECE, PNEUMATIC
Manufacturer (Section D)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer (Section G)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer Contact
jennifer rapsavage
56 e. bell drive
warsaw, IN 46582
5745260384
MDR Report Key17766048
MDR Text Key323634736
Report Number0001526350-2023-01150
Device Sequence Number1
Product Code GFD
Combination Product (y/n)N
Reporter Country CodeAS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 10/06/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/18/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number00880100100
Device Lot Number62961786
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received10/06/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/19/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberN/A
Patient Sequence Number1
Patient SexPrefer Not To Disclose
-
-