Model Number N/A
Device Problems Material Frayed (1262); Defective Component (2292)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/31/2023
Event Type  malfunction  
Event Description
It was reported that the patient had the product installed due to the need for dialysis.When placing per the standard operating procedure, the doctor found that the guide wire was deformed.No other information was provided.
 
Manufacturer Narrative
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The device has not been returned to the manufacturer for evaluation.H3 other text : device not returned for evaluation.
 
Event Description
It was reported that the patient had the product installed due to the need for dialysis.When placing per the standard operating procedure, the doctor found that the guide wire was deformed.No other information was provided.
 
Manufacturer Narrative
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The complaint of damaged guidewire was confirmed.The product returned for evaluation was stainless-steel j-tip guidewire.The investigation findings are consistent with damage caused by retraction of the guidewire against the bevel of the introducer needle.The returned product sample was evaluated and the guidewire was confirmed to be broken which allowed the outer coil wire to become unraveled.Microscopic examination of the fracture sites revealed the following: ¿ narrowing of the wire cross-section near the fracture site, which is a characteristic feature of a strong pull on the wire ¿ damage to the inside edge of the introducer needle which can occur if the guidewire is forcefully retracted against the needle, damaging the shape of the sharpened bevel biological material was also seen on the wire which may have contributed to the observed guidewire fracture as retraction of the wire together with the biological material could have caused the pieces to become stuck.Normal movement of the guidewire is away from the sharpened bevel and will not damage the wire.If the guidewire direction is reversed, the guidewire is then pulled against the sharpened edge of the needle bevel and can cause shearing damage of the wire and/or cause the wire to become stuck within the needle.An examination of the wire structure revealed no potential damage/defect related to manufacture of the product.H3 other text : evaluation findings are in section h.11.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NIAGARA SLIM-CATH DIALYSIS CATHETER KIT 12F X 20CM
Type of Device
CATHETER, HEMODIALYSIS, NON-IMPLANTED
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas 88780
MX   88780
Manufacturer Contact
kayla bevins
605 north 5600 west
salt lake city 84116
8015950700
MDR Report Key17769316
MDR Text Key324007588
Report Number3006260740-2023-04161
Device Sequence Number1
Product Code MPB
UDI-Device Identifier00801741045363
UDI-Public(01)00801741045363
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K010778
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 02/20/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/18/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number5553200
Device Lot NumberREGQ1654
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/20/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/01/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1