Model Number G151
Device Problems Incorrect, Inadequate or Imprecise Result or Readings (1535); Delayed Charge Time (2586); Battery Problem (2885)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/01/2022
Event Type  Injury  
Event Description
It was reported that this cardiac resynchronization therapy defibrillator (crt-d) had triggered code 1007 due to capacitor charge time exceeding limitations.The device had reached elective replacement indicator (eri) a year back and currently at end of life (eol) status.Subsequently this device was explanted, and a new device was successfully implanted.No additional adverse patient adverse effects were reported.The device is expected to return for analysis.
 
Event Description
It was reported that this cardiac resynchronization therapy defibrillator (crt-d) had triggered code 1007 due to capacitor charge time exceeding limitations.The device had reached elective replacement indicator (eri) a year back and currently at end of life (eol) status.Subsequently this device was explanted, and a new device was successfully implanted.No additional adverse patient adverse effects were reported.The device is expected to return for analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DYNAGEN CRT-D
Type of Device
CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATORS (CRT-D)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key17772753
MDR Text Key323713079
Report Number2124215-2023-51069
Device Sequence Number1
Product Code NIK
UDI-Device Identifier00802526534638
UDI-Public00802526534638
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010012/S341
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 09/29/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/19/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date01/29/2018
Device Model NumberG151
Device Catalogue NumberG151
Device Lot Number114442
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received09/28/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/11/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention; Hospitalization;
Patient Age74 YR
Patient SexFemale
-
-