• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION HOMECHOICE; SYSTEM, PERITONEAL, AUTOMATIC DELIVERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION HOMECHOICE; SYSTEM, PERITONEAL, AUTOMATIC DELIVERY Back to Search Results
Catalog Number 5C4471
Device Problem Unintended Electrical Shock (4018)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/22/2023
Event Type  malfunction  
Manufacturer Narrative
H10: the device was received for evaluation.During internal and external visual inspection, the elbow fitting, valve manifold (3 way), inner door gasket and assy cable digital were observed damaged.The device was powered on and the reported issue was not duplicated.The reported condition was not verified.The damaged components did not contribute with the reported issue or in connection with the reported problem.The damaged components were replaced.A service history review revealed no indication that the parts replaced during servicing caused or contributed to the reported event.Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported during the dwell phase of peritoneal dialysis therapy with a homechoice automated pd system the patient felt a ¿small electric shock in the stomach¿ while laying down.The patient visited the doctor and was told ¿the patient was in good health¿.The device had been plugged directly into a grounded outlet with no power cord defect.The patient cleaned and disinfected the machine with "a solution of water and chlorine and dry cloth".The patient was not touching any other device with electrical voltage.No footwear was worn at the time of the event.No signs of burns were noted to the machine.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HOMECHOICE
Type of Device
SYSTEM, PERITONEAL, AUTOMATIC DELIVERY
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - SINGAPORE
2 woodlands industrial park d
singapore 73875 0
SN   738750
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key17777532
MDR Text Key323800513
Report Number1416980-2023-04725
Device Sequence Number1
Product Code FKX
Combination Product (y/n)N
Reporter Country CodeMX
PMA/PMN Number
K102936
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 09/19/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/19/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Catalogue Number5C4471
Device Lot Number672465
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/04/2023
Date Manufacturer Received08/22/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-