WILLIAM A. COOK AUSTRALIA, PTY LTD OVUM ASPIRATION NEEDLE SINGLE LUMEN; MQE NEEDLE, ASSISTED REPRODUCTION - OVUM PICK-UP ASPIRATION NEEDLES Back to Search Results
Catalog Number K-OPS-6035-RWH-B-ET
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Hemorrhage/Bleeding (1888)
Event Date 07/18/2023
Event Type  Injury  
Manufacturer Narrative
(b)(4) are all from the same hospital.All were reported by the customer to the french regulatory authority who informed the manufacturer of the occurrences.
 
Event Description
The patient had an oocyte puncture echo-guided during ivf treatment.There was no anomaly on the device during the procedure.In the evening, emergency coelioscopy was done for drainage of hemoperitoneum in another hospital.
 
Event Description
The patient had an oocyte puncture echo-guided during ivf treatment.There was no anomaly on the device during the procedure.In the evening, emergency coelioscopy was done for drainage of hemoperitoneum in another hospital.
 
Manufacturer Narrative
No part of the complaint device relating to this event returned for evaluation.No imaging was received to assist the investigation.Review of device history record (dhr) found the work order for lot a1119110 and the related room stock work orders appeared complete, and the quality control inspection was verified to ensure that the devices passed inspection.One non-conforming device was rejected due to bent needle.There is no evidence that a device non-conformance or deficiency contributed to the reported event.The associated inspection record confirms that the devices were manufactured to specification.There were no temporary deviations in place at the time of manufacture.Review of specifications found that there are a number of processes and checks in place which would identify a blunt or damaged needle prior to shipment.The medical advisor reviewed the reported event and stated: "haemoperitoneum¿ by definition, this is the presence of blood lying freely within the peritoneal cavity.In most cases it implies here a hole in a blood vessel that cannot seal itself over by the usual clotting, spasm and wound healing.This is a dangerous, rare, well recognized complication of oocyte retrieval, in which sharp needles puncture ovarian capsules, then follicles which are highly metabolically active, with lots of blood flow through the walls.It is generally taught that the incidence of this complication is 1/1,000 egg retrievals (or slightly less).It is possible we are just seeing a statistical ¿clump¿ of cases.Haemoperitoneum was the case in all the laparoscopies and coelioscopy ¿ with blood losses around 1-2 liters.Once again, all these cases seemed to happen in french ivf units.I have no idea why that should be so, though it does suggest a systemic cause may have an influence".Clinical evaluation report for ovum pick-up needles and sets addresses the following: hemorrhage/bleeding is well-known and described in the literature as a possible adverse event occurring from the use of any ovum pick-up needle during the oocyte collection procedure.In conclusion, the cook ovum pick-up needles and sets are considered to be safe and effective and to be in accordance with the state-of-the-art for their intended use.An earlier corrective and preventative action (capa) was initiated to address an increase of bleeding after using cook ovum pick-up needles.After an extensive investigation into manufacturing processes, temporary deviations during manufacture from jan-2019 to jun-2022, needle design, design changes on single and double lumen devices over the last 3 years, assessment of clinical factors, and evaluation of returned complaint products both in-house and by puncture strength, puncture durability and drag force test, a single root cause could not be identified.The manufacturer has a quality management system (qms) requirement to always review the complaint history during a complaint investigation to ensure that trends are constantly monitored.Based on the available information, and inspection of the returned product, a definitive root cause could not be determined.The potential root causes are: patient related factors; procedural complications.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OVUM ASPIRATION NEEDLE SINGLE LUMEN
Type of Device
MQE NEEDLE, ASSISTED REPRODUCTION - OVUM PICK-UP ASPIRATION NEEDLES
Manufacturer (Section D)
WILLIAM A. COOK AUSTRALIA, PTY LTD
95 brandl street
eight mile plains
brisbane
AS 
Manufacturer (Section G)
WILLIAM A. COOK AUSTRALIA, PTY LTD
95 brandl street
eight mile plains
brisbane
AS  
Manufacturer Contact
tia gibb
95 brandl street
eight mile plains
brisbane 
AS  
MDR Report Key17778043
MDR Text Key323786445
Report Number9680654-2023-00134
Device Sequence Number1
Product Code MQE
UDI-Device Identifier00827002209435
UDI-Public(01)00827002209435(17)251004(10)A1119110
Combination Product (y/n)N
PMA/PMN Number
K983593
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Distributor
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 10/24/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/19/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue NumberK-OPS-6035-RWH-B-ET
Device Lot NumberA1119110
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received09/14/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/04/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age34 YR
Patient SexFemale
-
-