• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON DICKINSON & CO. (SPARKS) BD MAX¿ STAPHSR; SYSTEM, NUCLEIC ACID AMPLIFICATION TEST, DNA, METHICILLIN RESISTANT STAPHYLOCOCC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON DICKINSON & CO. (SPARKS) BD MAX¿ STAPHSR; SYSTEM, NUCLEIC ACID AMPLIFICATION TEST, DNA, METHICILLIN RESISTANT STAPHYLOCOCC Back to Search Results
Catalog Number 443419
Device Problem False Negative Result (1225)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/24/2023
Event Type  malfunction  
Event Description
It was reported that while using bd max¿ staphsr that there was a false negative.The following information was provided by the initial reporter: 3 samples that channels nuc (530/565 nm) and meca/mecc (585/630 nm) were positive, but mrej (475/520 nm) remained negative.
 
Manufacturer Narrative
E.1 initial reporter e-mail: (b)(6).H.3.A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event Description
It was reported that while using bd max¿ staph that there was a false negative.The following information was provided by the initial reporter: 3 samples that channels nuc (530/565 nm) and meca/mecc (585/630 nm) were positive, but mrej (475/520 nm) remained negative.
 
Manufacturer Narrative
After further evaluation of the complaint, it has been determined that the previously submitted report 1119779-2023-01018 was sent in error.There was no report of serious injury, medical intervention, or reportable device malfunction.Therefore, this is not considered to be a reportable malfunction.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD MAX¿ STAPHSR
Type of Device
SYSTEM, NUCLEIC ACID AMPLIFICATION TEST, DNA, METHICILLIN RESISTANT STAPHYLOCOCC
Manufacturer (Section D)
BECTON DICKINSON & CO. (SPARKS)
7 loveton circle
sparks MD 21152
Manufacturer (Section G)
BECTON DICKINSON & CO. (SPARKS)
7 loveton circle
sparks MD 21152
Manufacturer Contact
jennifer suh
5859 farinon drive
san antonio, TX 78249
8448235433
MDR Report Key17778330
MDR Text Key323813031
Report Number1119779-2023-01018
Device Sequence Number1
Product Code NQX
Combination Product (y/n)N
Reporter Country CodeSZ
PMA/PMN Number
K132822
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 11/09/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/19/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date03/11/2024
Device Catalogue Number443419
Device Lot Number2270650
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/28/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/27/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-