• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OTICON MEDICAL - NEURELEC NEURO ZTI COCHLEAR IMPLANT; NEURO ZTI EVO

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OTICON MEDICAL - NEURELEC NEURO ZTI COCHLEAR IMPLANT; NEURO ZTI EVO Back to Search Results
Model Number NEURO ZTI EVO
Device Problems Electrical /Electronic Property Problem (1198); Defibrillation/Stimulation Problem (1573)
Patient Problem Undesired Nerve Stimulation (1980)
Event Date 08/23/2023
Event Type  malfunction  
Manufacturer Narrative
While this device has gained initial pma approval, it has not been commercialized, imported or sold in the us and no implantations have been performed as of the date of this report.The implant neuro zti evo was not explanted as of the date of this report.Currently, this event is not a serious injury.Follow-up report will be submitted once the device explanted.The subject device is part of the voluntary field corrective action initiated for neuro zti on october 14th 2021 (international recall #: 211014).
 
Event Description
The oticon medical reprensentative reported : "during troubleshooting session at clinic in (b)(6), poland: patient reports unable to use implant since april due to sudden onset of painful shocking sensation and ipsilateral facial nerve stimulation (fns) when placing sound processor on head.External equipment was exchanged but did not resolve the problem.On (b)(6) 2023, using clinic sound processor, unable to run impedances without patient immediately removing sound processor due to painful shocking sensation and fns.Since three different sets of equipment have produced the same result, additional in-vivo troubleshooting is not warranted given the painful symptoms produced.The implant is clearly not operating within specifications and should be replaced under warranty." at this stage, the date of the scheduled explantation was not communicated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NEURO ZTI COCHLEAR IMPLANT
Type of Device
NEURO ZTI EVO
Manufacturer (Section D)
OTICON MEDICAL - NEURELEC
2720, chemin saint-bernard
porte 14
vallauris, 06220
FR  06220
Manufacturer (Section G)
OTICON MEDICAL - NEURELEC
2720, chemin saint-bernard
porte 14
vallauris, 06220
FR   06220
Manufacturer Contact
adrian ternisien
2720, chemin saint-bernard
porte 14
vallauris, 06220
FR   06220
MDR Report Key17779372
MDR Text Key323799689
Report Number3016743107-2023-00022
Device Sequence Number1
Product Code MCM
Combination Product (y/n)N
Reporter Country CodePL
PMA/PMN Number
P200021
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 09/20/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/20/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date11/09/2023
Device Model NumberNEURO ZTI EVO
Device Catalogue NumberM80185
Device Lot Number20-04875
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/23/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/23/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-