While this device has gained initial pma approval, it has not been commercialized, imported or sold in the us and no implantations have been performed as of the date of this report.The implant neuro zti evo was not explanted as of the date of this report.Currently, this event is not a serious injury.Follow-up report will be submitted once the device explanted.The subject device is part of the voluntary field corrective action initiated for neuro zti on october 14th 2021 (international recall #: 211014).
|
The oticon medical reprensentative reported : "during troubleshooting session at clinic in (b)(6), poland: patient reports unable to use implant since april due to sudden onset of painful shocking sensation and ipsilateral facial nerve stimulation (fns) when placing sound processor on head.External equipment was exchanged but did not resolve the problem.On (b)(6) 2023, using clinic sound processor, unable to run impedances without patient immediately removing sound processor due to painful shocking sensation and fns.Since three different sets of equipment have produced the same result, additional in-vivo troubleshooting is not warranted given the painful symptoms produced.The implant is clearly not operating within specifications and should be replaced under warranty." at this stage, the date of the scheduled explantation was not communicated.
|