Catalog Number 5000-01-01 |
Device Problem
Disconnection (1171)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 09/11/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
|
Event Description
|
It was reported that there was a temperature problem in an arctic sun device that was not cooling under 24°c.No water in chiller tank, chiller temperature (t4) was 24 ° c.The connector of the mixing pump was disconnected.Per review of wo on 18sep2023, there was no patient involvement.
|
|
Manufacturer Narrative
|
Upon further review, bd has determined that this mdr was reported in error as it was found to be a duplicate of an event previously reported.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
|
Event Description
|
It was reported that there was a temperature problem in an arctic sun device that was not cooling under 24°c.No water in chiller tank, chiller temperature (t4) was 24 ° c.The connector of the mixing pump was disconnected.Per review of wo on 18sep2023, there was no patient involvement.
|
|