OAKDALE LEVEL 1 HOTLINE FLUID WARMER, 390; WARMER, THERMAL, INFUSION FLUID Back to Search Results
Model Number HL-390
Device Problems Display Difficult to Read (1181); Poor Quality Image (1408)
Patient Problem Insufficient Information (4580)
Event Type  malfunction  
Manufacturer Narrative
Other, other text: b3: date of event and d4: udi section are unknown, no information has been provided to date.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
 
Event Description
It was reported that the fluid warmer display was dim, not readable.Patient involvement is unknown.
 
Manufacturer Narrative
Device evaluation: one device was retuned for investigation.Visual inspection noted: device received with an enclosure that had dents in it, front cover dented, water tank cover cracked, lcd broken, quick connect corroded, line cord faded and worn, and the microswitch was bent.Inspectors could visually observe that the lcd was unreadable due to damage.Functional testing found the pump was not circulating, confirming the customer complaint.The cracked lcd display showed ink leakage from impact damage.Root cause was attributed to customer impact damage.The product's history records were reviewed and there were no non-conformances nor service-related issues that would have resulted in the reported complaint.No action taken due to the condition of the device.It is deemed beyond economical repair and will be scrapped.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEVEL 1 HOTLINE FLUID WARMER, 390
Type of Device
WARMER, THERMAL, INFUSION FLUID
Manufacturer (Section D)
OAKDALE
3350 granada ave n
oakdale MN 55128
Manufacturer (Section G)
NULL
Manufacturer Contact
jim vegel
MDR Report Key17783079
MDR Text Key323859775
Report Number3012307300-2023-09135
Device Sequence Number1
Product Code LGZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K911383
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 11/15/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/20/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberHL-390
Device Catalogue NumberCON-HL-390
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received10/27/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/01/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-