BAXTER HEALTHCARE CORPORATION CLEARLINK CONTINU-FLO SOLUTION SET; SET, ADMINISTRATION, INTRAVASCULAR Back to Search Results
Catalog Number 2H8537
Device Problem Incomplete or Inadequate Connection (4037)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/26/2023
Event Type  malfunction  
Manufacturer Narrative
E1: additional initial reporter phone no.: (b)(6).G1: device manufacturer address 1: (b)(6).Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that a clearlink system non-dehp continu-flo solution set was unable to connect as the connector was unable to retract.This was discovered during set-up and preparation.There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
H10: the device was received for evaluation.A visual inspection was performed which observed the collar was bonded to the male luer which affected the mobility or spin of the collar.The reported condition was verified.The cause of the condition was improper solvent application during the automated assembly of the device's tail end segment during the manufacturing process.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLEARLINK CONTINU-FLO SOLUTION SET
Type of Device
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - CARTAGO
see h10
see h10
cartago 30106
CS   30106
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key17785267
MDR Text Key323876432
Report Number1416980-2023-04761
Device Sequence Number1
Product Code FPA
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
K961225
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 10/26/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/20/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number2H8537
Device Lot NumberR23A02031
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/23/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/02/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-