It was reported that the attachment device either seizes up and stops working or black debris rolls out of the saw head upon use.It was reported that upon evaluation the debris was inside the unit.It was reported that the once the gross debris is removed , the saw head is soaked in an enzymatic solution and then the saw is run between the warm water until debris can no longer be seen coming out.It was not reported if the device was used in surgery.There was no human patient involvement as the device was for veterinary use.The exact date of this event was unknown.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: as of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.Udi:(b)(4).
|