SYNTHES GMBH 1.5MM LOCKING SCREW SLF-TPNG WITH T4 STARDRIVE RECESS 6MM; PLATE, FIXATION, BONE
|
Back to Search Results |
|
Catalog Number 02.214.006 |
Device Problems
Improper or Incorrect Procedure or Method (2017); Difficult to Advance (2920)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 09/06/2023 |
Event Type
malfunction
|
Event Description
|
Device report from synthes reports an event in japan as follows: it was reported that on (b)(6) 2023, the patient underwent an unknown surgery.When inserting the locking screw, the surgeon was able to insert it halfway, but it turned idly, and the locking screw did not advance all the way to the contralateral side.It was noted that if another screw of the same length was used, that screw was inserted without any problems.The surgery was completed successfully with no delay.It was later reported that the cause of the event was an error in the surgical technique.This report involves one 1.5mm locking screw slf-tpng with t4 stardrive recess 6mm.This is report 1 of 1 for (b)(4).
|
|
Manufacturer Narrative
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: d9: complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.E3: reporter is a j&j employee.H3, h4, h6: without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Manufacturer Narrative
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Device used in a veterinary case - no patient information will be reported.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Search Alerts/Recalls
|
|
|