Model Number 412019
Device Problem Difficult to Remove (1528)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/30/2023
Event Type  malfunction  
Manufacturer Narrative
E1: name of affiliation: (b)(6).H6: medical device problem code: as per the information provided by complainant, for the issue dressing too sticky, the imdrf med.Dev.Prob.Code a050901 (adhesive too strong) is not available for selection.Based on the available information, this event is deemed to be a reportable malfunction.To date no additional information has been received.Should additional information become available, a follow-up report will be submitted.Fda registration number: reporting site: 1049092.Manufacturing site: 9618003.
 
Event Description
It was reported by a nurse that the dressing was really sticky to the patients¿ skin, and it was really difficult to remove.Also, reported that they did not use cleansers / skin protectors, just saline to clean the incision.The dressing was applied at the ward after twenty-four hours of the intervention and the dressing was used for four to six days.For removal, dressing was pulled from one corner and tried to take it out.No photograph was available at this time.
 
Manufacturer Narrative
Additional information - this emdr is being submitted to include the below: h6: investigation results under imdrf cause investigation code, imdrf investigation findings, imdrf cause conclusions h10: investigation summary photograph, video and/or physical sample evaluation: no photographs associated with this case were received and no unused return sample was expected.Batch record revision results: lot 3a00869 was manufactured on 08/jan/2023, in scd line, with a total of (b)(4) market units (mkus).The complaint investigator performed a batch record review on 25/sep/2023, to verify if all the applicable procedures were followed and no issues were found; all the components for assembly were correct per bill of materials (bom) and all the tooling information documented was also correct, system application product (sap) material id 1257951 and manufacturing order (b)(4).The production process, in-process control, testing results and packaging of products was run according to the applicable process instruction - process instructions for manual assembly and packaging - scd.Review of the batch records showed that all relevant tests required during the manufacturing process and final product release had been fulfilled and met the requirements.No discrepancy related to this issue were found within the documentation.Historical complaints review: on 25/sep/2023, complaint investigator ran a query in database from 01/jan/2022 to 25/sep/2023 in order to verify the complaints reported for the lot number 3a00869 for the malfunction code ¿adhesive dressing requires excessive force for removal from skin (i.E., difficult to remove)¿ and as result, no additional complaints were found.Historical nonconformance review: on 25/sep/2023, complaint investigator ran a query in database looking for any in process nonconformance / corrective action/preventive action (capa) (s) associated to the malfunction code ¿adhesive dressing requires excessive force for removal from skin (i.E.Difficult to remove)¿ for the lot number 3a00869 and as result, no nonconformance / corrective action/preventive action (capa) (s) for this malfunction code were generated during the manufacturing process of the referenced lot.Defect rate analysis: there has only been 8 defective parts confirmed to date from a lot size of 23,400 products.This represents a defect rate of only 0.03% which is well within an appropriate acceptable quality level (aql) for difficult to remove - difficult to remove issues which should be 0.65% based on our standard operating procedure (sop).In addition, all of the in-process testing on this lot did not find an adhesive issues, which confirms that the lot is unlikely to breach an acceptable quality level (aql) of 0.65.This issue certainly appears to be an isolated incident, but more importantly to date, it is well within our accepted acceptable quality level (aql) level for this type of failure mode or defect.Conclusions: no objective evidence was received such as a photo or sample, for this reason, we cannot conclude that the product does not meet specifications.The review of the batch record showed that all relevant tests required during the manufacturing process and final product release had been fulfilled and met the requirements.No nonconformity has been registered during the manufacturing process of the affected lot for the malfunction code ¿dressing/ribbon/gel handles insufficient wound exudate or does not retain absorbed wound exudate¿.No additional complaints were reported for lot affected related to the malfunction code ¿adhesive dressing requires excessive force for removal from skin (i.E.Difficult to remove)¿.Based on this, no negative trend was identified.This issue will be monitored through the post market product monitoring review process.To date no additional information has been received.Should additional information become available, a follow-up report will be submitted.Fda registration number reporting site: 1049092 manufacturing site: 9618003.
 
Event Description
To date no additional patient or event details have been received.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AQUACEL SURGICAL
Type of Device
DRESSING,WOUND,HYDROPHILIC
Manufacturer (Section D)
CONVATEC DOMINICAN REPUBLIC INC
km 18.5 parque industrial
itabo, s.a. haina
san cristobal 91000
DR  91000
Manufacturer (Section G)
CONVATEC DOMINICAN REPUBLIC INC
km 18.5 parque industrial
itabo, s.a. haina
san cristobal 91000
DR   91000
Manufacturer Contact
pamela meadows
7815 national service road
suite 600
greensboro, NC 
3365424679
MDR Report Key17797152
MDR Text Key324008267
Report Number9618003-2023-06223
Device Sequence Number1
Product Code NAC
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 08/30/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/22/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number412019
Device Lot Number3A00869
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Manufacturer Received09/30/2023
Was Device Evaluated by Manufacturer? No
Date Device Manufactured01/08/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-