Model Number 1058T
Patient Problem Failure of Implant (1924)
Event Type  Injury  
Event Description
It was reported that this cardiac resynchronization therapy defibrillator (crt-d) system was explanted during scheduled generator change out procedure.The device was at elective replacement indicator (eri).There was a high capture threshold on the left ventricular (lv) lead.There was an out-of-range impedance measurement on the right atrial (ra) lead, which had been turned off.It was decided by the physician to extract the leads in favor of new leads.A new crt-d was successfully placed.This report reflects information received by fda in the form of a notification per 803.22 (b)(2).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO
Type of Device
DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO
Manufacturer (Section D)
ST. JUDE MEDICAL
MDR Report Key17800148
MDR Text Key324171253
Report NumberMW5146063
Device Sequence Number1
Product Code NIK
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Voluntary
Reporter Occupation Other
Type of Report Initial
Report Date 09/01/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/21/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model Number1058T
Was Device Available for Evaluation? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-