• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) TENDRIL STS; No Match

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) TENDRIL STS; No Match Back to Search Results
Model Number 2088TC/52
Device Problem Material Twisted/Bent (2981)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/04/2023
Event Type  malfunction  
Manufacturer Narrative
The reported event of insulation twisted was confirmed.A complete lead was returned in one piece for analysis.Analysis found the outer insulation was twisted consistent with procedural damage.X-ray examination found the over-torqued inner coil at the connector region.Electrical testing did not find any indication of conductor fractures or internal shorts.
 
Event Description
It was reported that the patient presented for a scheduled procedure.During the procedure, it was noted that the insulation of the lead was twisted.The lead was not used, and a new lead was implanted.The patient was stable post-procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TENDRIL STS
Type of Device
No Match
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key17804747
MDR Text Key324132127
Report Number2017865-2023-46780
Device Sequence Number1
Product Code NVN
Combination Product (y/n)Y
Reporter Country CodeCH
PMA/PMN Number
P960013
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 09/22/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/22/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number2088TC/52
Device Lot NumberP000170733
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/11/2023
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received09/04/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/22/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-