Model Number MMT-105NNBLNA |
Device Problems
Insufficient Flow or Under Infusion (2182); Patient Device Interaction Problem (4001)
|
Patient Problems
Fatigue (1849); Headache (1880); Hyperglycemia (1905); Nausea (1970); Malaise (2359); Polydipsia (2604)
|
Event Date 08/29/2023 |
Event Type
Injury
|
Event Description
|
Information received by medtronic indicated that the customer reported that inpen was not delivering insulin and experienced hyperglycemia with a blood glucose value of 500 mg/dl.The customer also reported that the screw movement issue.Troubleshooting was performed by the customer.No further patient complications were reported.The customer will discontinue the use of the inpen and will be returned for analysis.
|
|
Manufacturer Narrative
|
Currently it is unknown whether or not the device may have caused or contributed to the event.The device has been returned, but not yet evaluated.Further information will follow once the analysis has been completed.No conclusion can be drawn at this time.Medtronic, inc.(medtronic) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information in the time allotted and has provided as much information as is available to the company as of the submission date this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the fda 3500a form and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.Medtronic objects to the use of these words and others like it because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|
|
Manufacturer Narrative
|
Per visual inspection: no physical damage to cartridge holder or inpen front and back shell was noted.Cartridge holder locks properly in place.Inpen paired to the commercial app.Inpen received with leadscrew 3/4 of travel.Performed dust/debris investigation and found visible horizontal abrasions and sticky fluid on the outside of electronics housing were noted.This indicated debris was lodged between the dose knob and electronics housing causing the leadscrew anomaly.Unable to perform baseline/wireless functionality and displacement dose accuracy test.Inpen failed front cap investigation.Inpen front shell does not fit securely onto cartridge holder due to snap arm being cracked / broken.In conclusion: deconstructive analysis demonstrated visible horizontal abrasions on the electronics housing were noted.This indicated debris was lodged between the knob and electronics housing causing the leadscrew to retract.This can affect insulin delivery.Therefore, the customer concern of leadscrew anomaly was confirmed.Cap anomaly was confirmed.Unable to confirm high bg.Unable to confirm under delivery.Medtronic, inc.(medtronic) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information in the time allotted and has provided as much information as is available to the company as of the submission date this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the fda 3500a form and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.Medtronic objects to the use of these words and others like it because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|
|
Search Alerts/Recalls
|
|
|