The olympus representative reported on behalf of the customer that during the colonoscopy with lesion removal and after the ligation, the endoloop on the single use ligating device deployed correctly however, it remained attached to the spring.The emergency loop cutters were used to cut the string externally and the scope was removed over the endoloop catheter.The colonoscope was then reinserted and the loop cutter was used again to cut the endoloop from the spring/catheter, which was later completely removed.The user was able to use the emergency cutter to detach the loop and the procedure was completed.The loop stayed with the patient and eventually passed.The procedure lasted 79 minutes and was not prolonged.The anesthesia/sedation was extended approximately five to seven minutes to troubleshoot the device.The condition of the patient was not impacted by the failure.No further medical intervention was required as a result of the reported problem.
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.The subject device was manufactured in october 2022, but the specific date is unknown.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation and past investigation results, it is likely the reported event occurred due to one of the following mechanisms: mechanism 1: 1) the loop was surrounding the body tissue, and it was temporarily ligated by pulling the slider.2) the tube sheath was pushed out, and the distal end of the coil sheath went into the tube sheath.3) an attempt was made to detach the loop in state of above description 2).Therefore, the loop detached from the hook in the tube.4) while the hook was extending from the coil sheath, the loop moved towards the proximal side and went into the coil sheath.5) the hook was pulled.This caused the hook and the loop to retract into the coil sheath together.As a result, the loop and the hook got stuck inside the coil and could not move.6) since the loop and the hook got stuck together inside the coil, the loop did not detach when the slider was pulled.Mechanism 2: 1) the sheath was bent near the handle.2) the operating wire could not move because sliding resistance between the sheath and the operating wire increased.3) the operating pipe deformed or broke because the slider was forcefully operated.4) due to above, the loop could not detach from the hook.The event can be prevented by following the instructions for use which state: "do not strike or crush the coil sheath during operation.Doing so can damage the distal end of the coil sheath, which could make it impossible to detach the loop after ligation.In this case, ¿emergency treatment¿ and as shown ¿equipment to be used in an emergency¿." "do not remove the loop from the hook while the coil sheath is not extended from the tube sheath.Otherwise, the loop may be tangled with the hook and become impossible to be removed.In this case, ¿emergency treatment¿ and as shown ¿equipment to be used in an emergency." "do not hold the loop with the distal end of the tube sheath while the loop is surrounding the tissue.Otherwise, when the tissue is ligated, the loop may be detached from the hook in the tube sheath and tangled with the hook.That may make the loop impossible to be removed.In this case, ¿emergency treatment¿ and as shown ¿equipment to be used in an emergency." "never use excessive force to operate the instrument.This could damage the instrument." olympus will continue to monitor field performance for this device.
|