Model Number MAJ-1430
Device Problem Connection Problem (2900)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/30/2023
Event Type  malfunction  
Manufacturer Narrative
The subject device was returned to olympus for evaluation.During inspection and testing, the allegation was confirmed.There was a flickering image due to a rusted bad connector.In addition, the round connector was worn out.The investigation is ongoing and follow up with the user facility is currently being performed.A supplemental report will be submitted upon completion of the investigation or if any additional information is provided by the user facility.
 
Event Description
A customer reported to olympus when the videoscope cable exera ii is plugged in, the blade side is not reading.The event occurred during preparation for use.There was no report of patient harm.
 
Manufacturer Narrative
This supplemental report is to inform that upon further review, this is not a reportable malfunction.The initial medwatch reported the returned device had connector issues with the blade side was not reading when plugged in.However; per the legal manufacturer, this issue is not likely to cause or contribute to death or serious injury if the malfunction were to recur.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VIDEOSCOPE CABLE EXERA II
Type of Device
SCOPE CABLE
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer (Section G)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key17819708
MDR Text Key324289770
Report Number3002808148-2023-10405
Device Sequence Number1
Product Code FDF
UDI-Device Identifier04953170214509
UDI-Public04953170214509
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K051645
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 12/14/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/26/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMAJ-1430
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/01/2023
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received12/12/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/24/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-