Model Number HL-90
Device Problems Defective Alarm (1014); Failure to Calibrate (2440)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
It was reported that the device will not calibrate and continues alarming.No patient involvement reported.
 
Manufacturer Narrative
Other text: b3: date of event and d4: udi section are unknown, no information has been provided to date.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
 
Manufacturer Narrative
H3.Device evaluated by manufacturer and h6.Evaluation codes: updated.One device was received.Per visual inspection, damaged printed circuit board (pcb), bent micro switch, and "intermittent" membrane switch.Per functional testing, the device would not calibrate properly and keeps alarming confirming the customer complaint.The root cause was a damaged main pcb board and bent microswitch lever from usage.The product's history records were reviewed and there were no non-conformances nor service-related issues that would have resulted in the reported complaint.Replaced the pcb, membrane switch, and micro switch.Performed preventative maintenance (pm).The device passed all functional and delivery tests.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEVEL 1 HOTLINE FLUID WARMER
Type of Device
WARMER, THERMAL, INFUSION FLUID
Manufacturer (Section D)
OAKDALE
3350 granada ave n
oakdale MN 55128
Manufacturer (Section G)
NULL
Manufacturer Contact
jim vegel
MDR Report Key17828948
MDR Text Key324406880
Report Number3012307300-2023-09283
Device Sequence Number1
Product Code LGZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K001764
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 11/16/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/27/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberHL-90
Device Catalogue NumberCON-HL-90
Was Device Available for Evaluation? No
Date Returned to Manufacturer09/13/2023
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received10/27/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/11/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-