D4.Udi, d10.Device available for evaluation, h3.Device evaluated by manufacturer and h6.Evaluation codes: updated.D3, g1,2 email is: (b)(6).One device was received with the enclosure having some kind of black dye on the right side, front cover global display was dented, printed circuit board (pcb) missing a light emitting diode (led).Per functional testing, the pump was not circulating.The temperature would not rise.The display could not be calibrated, it stayed at 34.6.The complaint was confirmed.The root cause was a faulty pump and pcb.The product's history records were reviewed and there were no non-conformances nor service-related issues that would have resulted in the reported complaint.No action taken due to the condition of the device.It was deemed beyond economical repair and was scrapped.
|