BECTON DICKINSON, S.A. BD POSIFLUSH¿ PRE-FILLED SALINE SYRINGE; SALINE, VASCULAR ACCESS FLUSH
|
Back to Search Results |
|
Catalog Number 306575 |
Device Problems
Delivered as Unsterile Product (1421); Component Missing (2306)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 09/13/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
B.3.Date of event: unknown.The date received by manufacturer has been used for this field.H.3.A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
|
|
Event Description
|
It was reported that the bd posiflush¿ pre-filled saline syringe was missing the tip cap.The following was received by the initial reporter: found one pieces of the posiflush, as showed in the photo without cap.
|
|
Manufacturer Narrative
|
A device history record review was completed for provided material number 306575 and lot number 3039182.The review did not reveal any detected abnormalities during the production process that could have contributed to the reported defect and all quality tests were found to be within specification.To aid in the investigation of this issue, one (1) picture sample was received for evaluation by our quality team.Through examination of the picture, two (2) syringes were shown in their flow wrap packaging.One (1) of the syringes had no tip cap.This defect may have resulted from an improperly screwed tip cap which detached after the syringe passed through the vision camera system.Based on the preventive measures in place, we believe this was an isolated incident with an unlikely chance of recurrence.At this time, further action has not been determined necessary.Our quality team will continue to closely monitor the manufacturing process for signs of this potential defect and any emerging trends.
|
|
Search Alerts/Recalls
|
|
|