Model Number 55750017545 |
Device Problem
Loosening of Implant Not Related to Bone-Ingrowth (4002)
|
Patient Problems
Undesired Nerve Stimulation (1980); Hernia (2240)
|
Event Date 09/05/2023 |
Event Type
Injury
|
Manufacturer Narrative
|
A: patient information cannot be included in regulatory report due to regional privacy regulations.G2: the country of the event is japan.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|
|
Event Description
|
Information was received from healthcare provider (hcp) via a manufacturer representative regarding a patient having l5/s1 tlif.It was reported that the patient underwent tlif on (b)(6) 2023 at l5/s1, and the cy cage contracted in august causing the grafted bone to compress the nerve. the grafted bone was coming out like a hernia and pressing on the nerve, so the cage was removed and replaced with milestone in an additional surgery performed on (b)(6) 2023.The screws were loose and the procedure was completed by replacing the loose screws. there were no further complications reported regarding the event.
|
|
Manufacturer Narrative
|
Additional information: h10 radiographic image review: the saggital ct reconstruction shows interbody cage at l5-s1.Lateral x-ray shows l5-s1 fusion construct and interbody grafts.No obvious hardware complications identified.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
Additional information: d9, h3, h6 h3.Device evaluation summary: product analysis #705799504:part # 55750017545 lot # h5828610 visual and optical inspection revealed the top portion of the crown of the screw has been worn from the seating of the rod.Optical inspection did not reveal any damage to the threads.Functional inspection confirmed the screw head was able to pivot in all directions and a sample break off screw was able to be threaded in the tulip head without any issue.The screw appears to function as intended.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|