MAUDE Adverse Event Report: ALCON RESEARCH, LLC - HOUSTON CONSTELLATION SURGICAL PROCEDURE PACK; GENERAL SURGERY TRAY (KIT)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - HOUSTON CONSTELLATION SURGICAL PROCEDURE PACK; GENERAL SURGERY TRAY (KIT) Back to Search Results
Catalog Number 8065000096
Device Problem Difficult to Open or Close (2921)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/29/2023
Event Type  malfunction  
Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event Description
A physician reported that the trocar valve open or close function failure during surgery.The procedure was completed after the product was replaced with another pack.There was no harm to patient.
 
Manufacturer Narrative
Two opened trocar cannula hub assemblies and two trocar handle blade assemblies, in a plastic sleeve, in a tray with other items were received for the report.Samples were visually inspected and found to be nonconforming.Sample #1 and #4 - damaged septum/ angled slit observed.Sample #2 and #3 - opened septum/ fish mouth observed.Leak testing could not be performed due to the damage of the samples.A review of the device history record traceable to the reported lot number indicates that the product was processed and released according to the product¿s acceptance criteria.The sample #2 and #3 evaluation confirms an open septum.The open condition of the valve could have contributed to the reported event of trocar valve failure.This is a manufacturing issue related to the over-molding of trocar hub.Samples #1 and #4, the exact root cause for this complaint is unknown; the damaged condition of the valve could have contributed to the reported event; how and when the valve became damaged cannot be determined from the evaluation performed.An investigation is currently in progress in order to further investigate issues with the trocar open septum of sample #2 and sample #3.Samples #1 and #4 the exact root cause for this complaint is unknown therefore specific action for this complaint cannot be taken.An acceptance quality inspection is performed to ensure product meets release acceptance criteria.This inspection includes evaluation for damaged valves.Complaints are reviewed and monitored at regular intervals to evaluate effectiveness of actions taken.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONSTELLATION SURGICAL PROCEDURE PACK
Type of Device
GENERAL SURGERY TRAY (KIT)
Manufacturer (Section D)
ALCON RESEARCH, LLC - HOUSTON
9965 buffalo speedway
houston TX 77054
Manufacturer (Section G)
ALCON RESEARCH, LLC - HOUSTON
9965 buffalo speedway
houston TX 77054
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key17835839
MDR Text Key324501203
Report Number1644019-2023-01301
Device Sequence Number1
Product Code LRO
UDI-Device Identifier00380650000965
UDI-Public00380650000965
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K880961
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 12/28/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/28/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number8065000096
Device Lot Number15CMD1
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/15/2023
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/04/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/24/2023
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
CONSTELLATION VISION SYSTEM.
-
-