Model Number HL-90
Device Problem Failure to Power Up (1476)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
It was reported that the device was not powering on.The issue was discovered during preventative maintenance.No harm or adverse effects were reported.
 
Manufacturer Narrative
Other text: b3: date of event and d4: udi section are unknown, no information has been provided to date.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
 
Manufacturer Narrative
Device evaluation: one device was returned for investigation.Visual inspection noted: device received with a corroded quick connect, water tank cover cracked on all four corners, plate clip discolored, line cord faded and worn, pole clamp had gouges in the center of holder, front cover was scratched, and the reflux plug was missing.Functional testing confirmed the complaint when the device would not turn on.Root cause was attributed to a faulty printed circuit board (pcb).The product's history records were reviewed and there were no non-conformances nor service-related issues that would have resulted in the reported complaint.Replaced pcb, quick connect, water tank cover, plate clip, line cord, pole clamp, reflux plug, and front cover.Performed preventative maintenance.Changed o rings and reservoir gasket.Calibrated device.Device passed all functional testing after the completed repairs.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEVEL 1 HOTLINE FLUID WARMER
Type of Device
WARMER, THERMAL, INFUSION FLUID
Manufacturer (Section D)
OAKDALE
3350 granada ave n
oakdale MN 55128
Manufacturer (Section G)
NULL
Manufacturer Contact
jim vegel
MDR Report Key17837084
MDR Text Key324512808
Report Number3012307300-2023-09347
Device Sequence Number1
Product Code LGZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K001764
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 11/15/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/28/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberHL-90
Device Catalogue NumberCON-HL-90
Was Device Available for Evaluation? No
Date Returned to Manufacturer09/21/2023
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received10/27/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/14/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-