OLYMPUS WINTER & IBE GMBH RESECTION SHEATH, 8 MM, FOR 8.5 MM/26 FR. OUTER SHEATH, ABS; RESECTOSCOPE SHEATH Back to Search Results
Model Number A42011A
Device Problems Component Missing (2306); Detachment of Device or Device Component (2907)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/04/2023
Event Type  malfunction  
Manufacturer Narrative
(b)(6).G4 pma/510(k) number: k931994 / k931995.The device was returned and evaluated, and the customer¿s allegation of ceramic tip missing was confirmed.Device evaluation found that the ceramic tip was damaged.The additional evaluation findings are as follows: the sealing rings were damaged and missing.The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation or if any additional information is provided by the user facility.
 
Event Description
The customer reported to olympus, the resection sheath ceramic tip was missing.The issue was found during reprocessing.There were no reports of patient harm.
 
Manufacturer Narrative
This supplemental report is being submitted to provide additional information based on the legal manufacturer's final investigation.Corrected d10 and h6 medical device problem code based on information available at the time of the initial submission.A review of the device history record found no deviations that could have caused or contributed to the reported issue.The device was returned to olympus for evaluation and the customer's allegation was confirmed.Based on the results of the investigation, the insulation insert encountered thermally and/or mechanically induced damaged.Therefore, the root cause is due to wear and tear and/or improper handling by the customer.This issue is addressed in the instructions for use (ifu) which state: 4 before use: warning - infection control risk properly reprocess the product before first and each subsequent use following the instructions in this manual and in the system guide endoscopy.Improper and/or incomplete reprocessing can cause infection of the patient and/or medical personnel.4.1 inspection and testing: inspecting the product visually inspect the product.Make sure that it has: -- no corrosion -- no dents -- no scratches ceramic insulation at distal end: visually inspect the ceramic insulation at the sheath¿s distal end before each use.Do not use the instrument in case of damage (e.G.Cracks, fractures).Warning - risk of injury impact, fall, shock or similar stress can damage the ceramic insulation at the sheath¿s distal end.Damaged instruments can cause injuries to the patient and/or user.Do not use the instrument if damaged.Damaged product: if the product is damaged or does not function properly, contact an olympus representative or an authorized service center.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESECTION SHEATH, 8 MM, FOR 8.5 MM/26 FR. OUTER SHEATH, ABS
Type of Device
RESECTOSCOPE SHEATH
Manufacturer (Section D)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg, hamburg 22045
GM  22045
Manufacturer (Section G)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg, hamburg
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key17841055
MDR Text Key324554792
Report Number9610773-2023-02757
Device Sequence Number1
Product Code HIH
UDI-Device Identifier04042761023658
UDI-Public04042761023658
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K931994
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 02/21/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/29/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberA42011A
Device Lot Number14XW-0008
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/07/2023
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received02/20/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/15/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
OTV-S190.
-
-