MAUDE Adverse Event Report: BIOMET UK LTD. OXF UNI TIB TRAY SZA LM; PROSTHESIS, KNEE, FEMOROTIBIAL, UNICOMPARTMENTAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOMET UK LTD. OXF UNI TIB TRAY SZA LM; PROSTHESIS, KNEE, FEMOROTIBIAL, UNICOMPARTMENTAL Back to Search Results
Model Number N/A
Device Problem Loosening of Implant Not Related to Bone-Ingrowth (4002)
Patient Problems Failure of Implant (1924); Pain (1994); Ambulation Difficulties (2544); Osteopenia/ Osteoporosis (2651); Swelling/ Edema (4577)
Event Date 09/07/2023
Event Type  Injury  
Manufacturer Narrative
(b)(4).D10 ¿ oxf anat brg lt sm size 4 pma; item#: 159541, lot#: 811040 oxf twin-peg cmntd fem sm pma; item#: 161468, lot#: 175150 unknown cement the device will not be returned for analysis; however, an investigation of the reported event is in progress.Once the investigation is completed, a supplemental medwatch 3500a will be submitted.
 
Event Description
It was reported that the patient underwent revision surgery to a total knee approximately one year and five months post initial left partial knee replacement, due to tibial baseplate loosening.Poor bone quality was noted during the surgery.Due diligence is in progress for this complaint; to date no additional information or product has been received.
 
Event Description
No further event information available at the time of this report.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.No product was returned or pictures provided; visual and dimensional evaluations could not be performed.A review of the device manufacturing records confirmed no abnormalities or deviations that could be related to the reported event.Device is used for treatment.Radiographs were provided and reviewed by a radiologist.Three views of the left knee demonstrate a medial compartment hemiarthroplasty with subsidence of the medial tibial component with radiolucency along the cement hardware interface consistent with loosening.Moderate joint effusion and osteopenia was also noted.With the available information, a definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OXF UNI TIB TRAY SZA LM
Type of Device
PROSTHESIS, KNEE, FEMOROTIBIAL, UNICOMPARTMENTAL
Manufacturer (Section D)
BIOMET UK LTD.
waterton industrial estates
bridgend CF31 3XA
UK  CF31 3XA
Manufacturer (Section G)
BIOMET UK LTD.
waterton industrial estates
bridgend CF31 3XA
UK   CF31 3XA
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key17842324
MDR Text Key324559584
Report Number3002806535-2023-00326
Device Sequence Number1
Product Code NRA
UDI-Device Identifier05019279388844
UDI-Public(01)05019279388844(17)300908(10)281270
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010014
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 10/17/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/29/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberN/A
Device Catalogue Number154718
Device Lot Number281270
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received10/13/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/08/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
Patient Outcome(s) Required Intervention; Hospitalization;
Patient SexFemale
Patient Weight71 KG
-
-