• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION ELUVIA DRUG-ELUTING VASCULAR STENT SYSTEM; STENT, SUPERFICIAL FEMORAL ARTERY, DRUG-ELUTING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION ELUVIA DRUG-ELUTING VASCULAR STENT SYSTEM; STENT, SUPERFICIAL FEMORAL ARTERY, DRUG-ELUTING Back to Search Results
Model Number 24657
Device Problem Activation Failure (3270)
Patient Problem Thrombosis/Thrombus (4440)
Event Date 08/07/2023
Event Type  Injury  
Manufacturer Narrative
A1 - patient identifier: (b)(6).A2 - age at time of event: the patient was 73 years old at the time of study enrollment.
 
Event Description
(b)(4).It was reported that the subject experienced stent thrombosis and stent under expansion during the index procedure.The subject underwent treatment with the eluvia drug-eluting stents and ranger drug-coated balloon on (b)(6) 2023 as a part of the (b)(4) clinical trial.The target lesion (001) was in the left proximal superficial femoral artery (sfa), left mid sfa, left distal sfa, left proximal popliteal artery, extending up to left mid popliteal artery.The target lesion had a 5.5 mm distal reference vessel diameter with 70% stenosis and was classified as tasc ii d lesion.Prior to the treatment of target lesion with the study device, pre-dilation was performed by using 4 mm x 220 mm and 6 mm x 120 mm sterling over the wire pta balloon.Treatment of target lesion was performed by dilation using study devices, placement of two 7 mm x 120 mm, one 7 mm x 60 mm, and one 7 mm x 80 mm eluvia drug-eluting stents, followed by 6 mm x 120 mm ranger drug-coated balloon angioplasty.Following which, post-treatment was performed with placement of a 7 mm x 250 mm non-boston scientific stent and balloon dilation was performed by using a 6 mm x 60 mm non-boston scientific balloon, 7 mm x 80 mm sterling over the wire pta balloon, and 7 mm mustang balloon.Post treatment, the final residual stenosis was noted to be 30%.During the treatment of target lesion (001), angiography performed revealed under expansion of the 7 mm x 120 mm eluvia drug eluting stent, placed in the left sfa.Following which, the stent was extended to endarterectomized portion of left sfa with 7 mm x 120 mm and 7 mm x 60 mm eluvia drug eluting stents in an overlapping manner.Subsequently, the under expanded stent was treated with balloon dilation using 7 mm x 80 mm sterling balloon, however, angiography revealed sluggish flow through sfa stent.Hence, attempt was made to expand the stent with lithotripsy using an unknown 6 mm x 60 mm balloon.Angiography performed revealed flow limiting dissection which was treated with placement of 7 mm x 80 mm eluvia stent.Post procedural angiography performed revealed under expanded eluvia drug eluting stent and possible thrombus in the stent placed in the left sfa.In response to the event, a non-boston scientific stent was deployed from above knee popliteal artery to the proximal sfa.Following which, aggressive angioplasty was performed to treat the under expanded stent with 7 mm mustang balloon which resulted in the full expansion of the sfa stent.Angiography performed revealed sluggish flow in the sfa stent with no evidence of residual stenosis or thrombus.On (b)(6) 2023, the subject was discharged from the hospital.
 
Manufacturer Narrative
A1 - patient identifier: (b)(6).A2 - age at time of event: the patient was 73 years old at the time of study enrollment.Updated fields: b5 - describe event or problem.B6 - relevant tests/laboratory data.B7 - other relevant history.
 
Event Description
(b)(4) elegance it was reported that the subject experienced stent thrombosis and stent under expansion during the index procedure.The subject underwent treatment with the eluvia drug-eluting stents and ranger drug-coated balloon on (b)(6) 2023 as a part of the elegance clinical trial.The target lesion (001) was in the left proximal superficial femoral artery (sfa), left mid sfa, left distal sfa, left proximal popliteal artery, extending up to left mid popliteal artery.The target lesion had a 5.5 mm distal reference vessel diameter with 70% stenosis and was classified as tasc ii d lesion.Prior to the treatment of target lesion with the study device, pre-dilation was performed by using 4 mm x 220 mm and 6 mm x 120 mm sterling over the wire pta balloon.Treatment of target lesion was performed by dilation using study devices, placement of two 7 mm x 120 mm, one 7 mm x 60 mm, and one 7 mm x 80 mm eluvia drug-eluting stents, followed by 6 mm x 120 mm ranger drug-coated balloon angioplasty.Following which, post-treatment was performed with placement of a 7 mm x 250 mm non-boston scientific stent and balloon dilation was performed by using a 6 mm x 60 mm non-boston scientific balloon, 7 mm x 80 mm sterling over the wire pta balloon, and 7 mm mustang balloon.Post treatment, the final residual stenosis was noted to be 30%.During the treatment of target lesion (001), angiography performed revealed under expansion of the 7 mm x 120 mm eluvia drug eluting stent, placed in the left sfa.Following which, the stent was extended to endarterectomized portion of left sfa with 7 mm x 120 mm and 7 mm x 60 mm eluvia drug eluting stents in an overlapping manner.Subsequently, the under expanded stent was treated with balloon dilation using 7 mm x 80 mm sterling balloon, however, angiography revealed sluggish flow through sfa stent.Hence, attempt was made to expand the stent with lithotripsy using an unknown 6 mm x 60 mm balloon.Angiography performed revealed flow limiting dissection which was treated with placement of 7 mm x 80 mm eluvia stent.Post procedural angiography performed revealed under expanded eluvia drug eluting stent and possible thrombus in the stent placed in the left sfa.In response to the event, a non-boston scientific stent was deployed from above knee popliteal artery to the proximal sfa.Following which, aggressive angioplasty was performed to treat the under expanded stent with 7 mm mustang balloon which resulted in the full expansion of the sfa stent.Angiography performed revealed sluggish flow in the sfa stent with no evidence of residual stenosis or thrombus.On (b)(6) 2023, the subject was discharged from the hospital.It was further reported that the target lesion had a proximal reference vessel diameter of 6.5 mm and a lesion length of 250 mm.On (b)(6) 2023 during the index procedure, occlusion noted in the left common femoral artery and sfa were treated with endarterectomy followed by pre-dilation of sfa and popliteal artery using 4mm x 220 mm sterling pta and angioplasty of left sfa with 6mm x 120 mm sterling balloon.Due to extensive calcific disease and multiple dissection planes, 7mm x 120 mm eluvia stent was placed from left distal sfa to proximal sfa.The previously reported flow-limiting dissection was further reported to be grade e in the left above knee popliteal artery, and was attributed to the 6mm x 120mm sterling balloon.Post procedure, angiography performed revealed filling defect with possible thrombus in the mid segment of left sfa stent.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELUVIA DRUG-ELUTING VASCULAR STENT SYSTEM
Type of Device
STENT, SUPERFICIAL FEMORAL ARTERY, DRUG-ELUTING
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
rachel shields
4100 hamline ave n
arden hills, MN 55112
6512422111
MDR Report Key17842992
MDR Text Key324577277
Report Number2124215-2023-53203
Device Sequence Number1
Product Code NIU
UDI-Device Identifier08714729876731
UDI-Public08714729876731
Combination Product (y/n)Y
Reporter Country CodeUS
PMA/PMN Number
P180011
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Study,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 12/20/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/29/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number24657
Device Catalogue Number24657
Device Lot Number0031481708
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/04/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/10/2023
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient SexMale
Patient RaceBlack Or African American
-
-