Catalog Number D134701 |
Device Problems
Partial Blockage (1065); Contamination /Decontamination Problem (2895)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 09/05/2023 |
Event Type
malfunction
|
Event Description
|
It was reported that a patient underwent an atrial fibrillation (afib) ablation procedure with a thermocool® smart touch® sf uni-directional navigation catheter which would not irrigate correctly and it was found to have a wire in the port.It was reported there was a wire or small tube in the valve for irrigation.The catheter was replaced and the procedure was continued.There was no patient consequence.
|
|
Manufacturer Narrative
|
On (b)(6) 2023, the bwi product analysis lab received the complaint device for evaluation.The product analysis has begun but is not completed at this time.When the investigational analysis has been completed, a supplemental 3500a report will be submitted.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by biosense webster inc., or its employees that the report constitutes an admission that the product, biosense webster inc., or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's ref.# (b)(4).
|
|
Manufacturer Narrative
|
It was reported that a patient underwent an atrial fibrillation (afib) ablation procedure with a thermocool® smart touch® sf uni-directional navigation catheter which would not irrigate correctly and it was found to have a wire in the port.It was reported there was a wire or small tube in the valve for irrigation.The catheter was replaced and the procedure was continued.There was no patient consequence.Device evaluation details: the device was returned to biosense webster inc (bwi) for evaluation and the evaluation has been completed.A visual inspection, and irrigation test was performed following bwi procedures.Visual analysis revealed no damage or anomalies on the device.The tube or wire reported by the customer could be related to the irrigation tube of the device which is part of the design of the catheter and no anomalies were found on it.An irrigation test was performed, and the device was irrigating correctly.No irrigation issues were found.A manufacturing record evaluation was performed for the finished device batch number, and no internal actions were identified.The issues reported by the customer could not be replicated during the product investigation; other issues or circumstances may have occurred during the usage of the device that compromised its performance.As part of biosense webster's quality process, all devices are manufactured, inspected, and released to approved specifications.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's ref.# (b)(4).
|
|
Search Alerts/Recalls
|
|